Cryogenic STM & SFM

Low Temperature 3He & 4He SPM Operation in High Magnetic Fields

Download this page as PDF

Technical Information

Cryogenic STM:

Scan range: XYZ = 6.5 µm x 6.5 µm x 0.8 µm at RT
XYZ = 1.2 µm x 1.2 µm x 0.15 µm at 4.2 K
Resolution: atomic resolution STM 2)
2D coarse positioning: Z: ~6 mm, X: ~4 mm

 

4 electrical sample contacts

In-situ exchangeable tip carriers with three electrical contacts

Cernox temperature sensor

Measurement modes: STM, STS, IETS, SP-STS, etc.

 

 

Cryogenic SFM:

Scan range: XYZ = 20 µm x 20 µm x 2.2 µm at RT
XYZ = 3.6 µm x 3.6 µm x 0.39 µm at 4.2 K
Resolution: atomic resolution STM and AFM2)
3D coarse positioning: X/Y: ~5 mm ø circle, Z: ~10 mm

 

10 electrical sample contacts

In-situ exchangeable tip carriers with four electrical contacts

Cernox temperature sensor

Measurement modes: NC-AFM, C-AFM, MFM, KPFM, STM, STS, IETS, SP-STS, etc.

 

Note: Specifications and descriptions contained in this website are subject to alteration without notice. For guaranteed specifications please inquire official documentation. 